Blog
Testing Report of Ozone Analyzer UV-2000S
O3tech
2014-6-22 14:23:02
112
O3 Tech (H.K) Limited | ||||||
BLDG C#, MEICHENG TECH PARK, CHANG ZHEN, GONGMING, GUANGMING NEW DISTRICT, SHENZHEN, CHINA 518000 | ||||||
Tel: 86.755.89969024 | ||||||
http://www.o3tech.cn; info@o3tech.cn | ||||||
Testing Report | ||||||
14-Apr-16 | ||||||
Ozone Analyzer:Ultraviolet Absorption. UV-2000S | ||||||
Cooling method: air cooling | ||||||
Temperature: 25 ℃ | |
Model | Ozone Tubes QTY | Voltage | Bottle Oxygen | O2 Flow/min | Ozone Concentration Mg/L | Ozone Output grams/H |
SP001G | 1 | 220V/HZ | 99% | 2 | 27.18 | 3.2616 |
1 | 34.78 | 2.0868 | ||||
1/4 | 45.56 | 0.6834 | ||||
1/8 | 43.82 | 0.32865 | ||||
1/16 | 42.68 | 0.16005 |
Ozone output calculating formula: | ||||||
Oxygen flow (LPM) ×60 Min(1 hour)× ozone concentration (MG/L)÷1000=Ozone output (Grams/Hour) | ||||||